Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20220
Brand Name:Medtronic Neuromodulation Lead Extensions
Model:7483 and 37086
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue: Post marketing surveillance for DBS therapy has identified the following serious events or injuries associated with tunneling that are not presented in the current DBS labeling. The following events are associated with the DBS implant and tunneling the extension from the lead to the INS:
"Cases of spinal accessory nerve injury
"Severe bleeding or vascular injury
"An extension inadvertently implanted through the ribs and later found to be close to the heart requiring revision
"Tunneling too superficially resulting in an extension that exited and then re-entered the neck which wasn't discovered until the surgical drape was removed
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:2/6/2016
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /