Detail of Recall

Type of Product:Device
Medsafe Reference:20193
Brand Name:Arjo-Huntleigh Concerto & Basic Shower Trolleys
Model:
Affected:batch of devices manufactured between 7th March 2014 and 20th April 2015.
Software version:
Recalling Organisation:Arjo New Zealand Limited
34 Vestey Drive
Mt Wellington
AUCKLAND 1060
Contact Information:0800 000 151
Manufacturer:ArjoHuntleigh AB
Issue:There have been a number of complaints where the bolts under the stretcher have become loose. There has been one minor injury as a result of a patient falling related to these incidents. Manufacturer investigated the issue and the conclusion was that he stretcher mounting bolts or stretcher tilting mechanism can become loose which may make the device unstable, providing a potential for risk to the patient or the caregiver. There is a risk the locking mechanism for the stretcher-tilt function may not engage properly, resulting in a patient falling to the floor.
The issue relates to a batch of devices manufactured between 7th March 2014 and 20th April 2015.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/5/2016