Detail of Recall

Type of Product:Device
Medsafe Reference:20188
Brand Name:Hospira Sapphire Infusion Pump
Model:L/N 163113601, L/N 163123601
Affected:
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Q Core Medical Ltd
Issue:In isolated cases false "Air-in-Line" alarms are generated when initiating an infusion of Iron sucrose at concentrations of 0.7mg/ml - 2.8 mg/ml.
The epidural pumps Hospira L/N 163123601/ Q Core P/N 15032-000-0001 use the same algorithm as in other Sapphire infusion pumps models therefore the epidural pumps will be upgraded when the new software is available.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/6/2016