Detail of Recall

Type of Product:Device
Medsafe Reference:20178
Brand Name:Dimension Vista System Enyme 2 Calibrator
Model:KC321
Affected:4MD048, 5DD026, 5ED071, 5GD004, 5JD58, 5LD061
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:There have been complaints received of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 Calibrator (certain lots). Confirmed that patients, quality control, and proficiency testing material all demonstrate a similar shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/5/2016