Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20167 |
| Brand Name: | Alcon Labs Accurus Consumables (IIb) - 45150 |
| Model: | 8065752192, 8065752200 |
| Affected: | 1794552H, 1799602H, 1799605H |
| Software version: | |
| Recalling Organisation: | Alcon Laboratories Australia Pty Ltd (NZ) c/- Pharmaco (NZ) Ltd 4 Fisher Crescent Mt Wellington Auckland |
| Contact Information: | 09 270 4360 |
| Manufacturer: | Alcon Laboratories Inc |
| Issue: | Due to the possible presence of a moulding irregularity found within the aspiration lue may result in a reduction of vacuum within the aspiration line. The affected lots may result in a small air leak and corresponding reduction in vacuum performance, leading to inability to achieve or reach maximum targeted vacuum and reduction of holding force at distal end of the phaco tip. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/5/2016 |