Detail of Recall

Type of Product:Device
Medsafe Reference:20166
Brand Name:LumaDent Battery Pack for Surgeon Head Light
Model:All model number afected except SW-084120
Affected:Part No: BP500v3.1
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:LumaDent Inc
Issue:The affected battery chargers have not been demonstrated to comply with the Australian and new Zealand Electrical Safety Standard thus may have the potential to overcharge the batteries causing them to over- heat and catch fire.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2016