Detail of Recall

Type of Product:Device
Medsafe Reference:20164
Brand Name:Smiths Medical CADD administration Set with flow stop
Model:CADD
Affected:Multiple Lots 21-7321-01
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical MD Inc
Issue:It has become aware that under delivery to patients of medicinal products may occur when administered using CADD® Administration Sets with Flow Stop. Test data indicates that under infusion has the potential to contribute to an average of an additional 5.2% under delivery beyond the +/- 6% stated in the Operator's Manual for the CADD® pumps.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/5/2016