Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20164 |
| Brand Name: | Smiths Medical CADD administration Set with flow stop |
| Model: | CADD |
| Affected: | Multiple Lots 21-7321-01 |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical MD Inc |
| Issue: | It has become aware that under delivery to patients of medicinal products may occur when administered using CADD® Administration Sets with Flow Stop. Test data indicates that under infusion has the potential to contribute to an average of an additional 5.2% under delivery beyond the +/- 6% stated in the Operator's Manual for the CADD® pumps. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/5/2016 |