Detail of Recall

Type of Product:Device
Medsafe Reference:20156
Brand Name:Rex Rehab
Model:REX Rehab
Affected:REX-2016-0082
Software version:
Recalling Organisation:Rex Bionics Limited
58 Apollo Drive
Rosedale
Auckland 0751
Contact Information:
Manufacturer:Rex Bionics Limited
Issue:As part of the corrective and preventive action into an incident report, where a patient sustained an injury relating to incorrectly securing the heel stops on the REX, the sponsor has issued a Recall for Product Correction to:
1. Reinforce the instructions in the current IFU around the importance of correct heel stop adjustment, and
2. Add an additional statement to the IFU to specifically highlight the use of external supports (such as braces, splints, supports) which may restrict a user's range of motion and cause serious harm.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/6/2016