Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20140
Brand Name:Phada AB EliA CCP Positive Control 250
Model:83-1-35-01
Affected:Lot Nos: C9VBF C9VBH and C9VBK (vials) HG56 HJPS HM8Z and J5N5 (Kits)
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:
Issue:When using the Phada EliA CCP Well (14-5515-01) in combination with EliA IgG Conjugate, anti-CCP-signals can be elevated for specific conjugate lots.
The positive bias of the EliA CCP test will change a small percentage of test results from negative to equivocal and equivocal or high negative to low positive.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/5/2016