Detail of Recall

Type of Product:Device
Medsafe Reference:20135
Brand Name:Siemens ADVIA Centaur, ADVIA Centaur XP,ADVIA Centaur XPT, ADVIA Centaur CP Tnl-Ultra
Model:Tnl-Ultra (100 test) 02789602, Tnl-Ultra (500) 02790309
Affected:multiple lot numbers
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that all in date lots of ADVIA Centaur TnI-Ultra exhibit a greater than 10% change in results in samples with biotin levels up to 10 ng/mL (41 nmol/L).
The Instructions for Use (IFU) state that specimens that have up to 10 ng/mL (41 nmol/L) of biotin demonstrate
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/5/2016