Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20125
Brand Name:Abbott CELL-DYN Emerald Cleaner
Model:09H46-02
Affected:6991, 7024, 7027
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd
Unit D
4 Pacific Rise
Mt Wellington
Auckland 2025
Contact Information:0800 656 233
Manufacturer:
Issue:Previously identified occurrences where the CELL-DYN Emerald analyzer generates Quality Control (QC) Out Of Range Low for parameters RBC and PLT when using Emerald Cleaner lots 6853, 6901, and 6953 (Medsafe Recall Reference #19975 )Further investigation identified additional lots 6991, 7024, and 7027 as having similar occurrences.
Recall action type:Correction
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/5/2016