Detail of Recall

Type of Product:Device
Medsafe Reference:20118
Brand Name:Hologic ThinPrep 5000 Processor with Autloloader (Instrument Analyser)
Model:
Affected:
Software version:
Recalling Organisation:Pharmaco (NZ) Ltd
Level 1
4 Fisher Crescent
Mt Wellington
AUCKLAND 1060
Contact Information:0800 804 079
Manufacturer:Hologic Inc
Issue:In rare cases, it is possible, for the ThinPrep® 5000 Processor with Autoloader to misread the value encoded in the vial barcode label. This occurs when the barcode scanner decodes a value that meets all symbology rules and check-digit validation but is not the intended value encoded in the barcode and processes the vial using this identifier.
In an Autoloader configuration, if a misread occurs, the misread vial ID value is transferred to the slide and a slide is produced with the incorrect identifier
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/5/2016