Detail of Recall

Type of Product:Device
Medsafe Reference:20116
Brand Name:Philips Brilliance 64 and Ingenuity CT
Model:Philips Brilliance 64, Ingenuity CT,Ingenuity Core, Ingenuity Core 128
Affected:software versions 4.1.3 and v4.1.4
Software version:4.1.3 and v4.1.4
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:Philips Brilliance iCT and Brilliance cCT SP
Model:Philips Brilliance iCT, Brilliant iCT SP
Affected:software versions 4.1.3 and v4.1.4
Software version:4.1.3 and 4.1.4
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Some issues have been detected in software version 4.1.3 and 4.1.4 in Philips Brilliance 64 and Ingenuity CT, Brilliance iCT, Brilliance iCT SP products .
When the scan length is set to a value that is very close to the whole number, the calculated reconstruction length is a value larger than the actual scan length. The scan increment changes when there is a change in Field of View (FOV) causing the scan length to change as well.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/5/2016