Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20111 |
| Brand Name: | Siemens ADVIA Centaur XPT System |
| Model: | 10711433 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | An issue has been identified with the ADVIA Centaur® XPT default setting for the Daily Maintenance Task (or Daily Cleaning Procedure) frequency. Instruments may have the Daily Maintenance Task frequency set to "As needed" instead of "Daily". Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/5/2016 |