Detail of Recall

Type of Product:Device
Medsafe Reference:20111
Brand Name:Siemens ADVIA Centaur XPT System
Model:10711433
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:An issue has been identified with the ADVIA Centaur® XPT default setting for the Daily Maintenance Task (or Daily Cleaning Procedure) frequency. Instruments may have the Daily Maintenance Task frequency set to "As needed" instead of "Daily". Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/5/2016