Detail of Recall

Type of Product:Device
Medsafe Reference:20083
Brand Name:Carestream - Image Suite
Model:
Affected:software versions 3.0 4.0
Software version:
Recalling Organisation:Carestream Health New Zealand Ltd
c/- Quigg Partners
Level 7, 28 Brandon St
Wellington
Contact Information:021 977 215
Manufacturer:Carestream Health Inc (HCIS)
Issue: A software issue related to image orientation. If the Image Suite software receives a Computed Tomography (CT )or magnetic resonance (MR) image with the image orientation (i.epatient) tag specified and the overlay is configured to display this tag, then the left side is marked R and the right marked L and the anterior marked P and the posterior marked A.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/4/2016