Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20080 |
| Brand Name: | Stryker Endoscopy CinchLock SS Knotless Anchor with Inserter |
| Model: | CAT02462 |
| Affected: | Lot Numbers: 15120801, 15121401, 16010402, 16011901, 16012501 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Endoscopy |
| Issue: | Manufacturer had advised that there have been several complaints related to the pull wire breaking since product launch.There were two modes of failure identified for the pull wire: 1. Near the inserter handle and 2.At the most distal tip where the wire ends with a larger diameter tip |
| Recall action type: | Correction |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/4/2016 |