Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20041
Brand Name:GE Medical CT Lightspeed Family
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Brand Name:GE Medical Nuclear Medicine System, Gamma Camera, Stationary
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Functional Imaging Haifa
Issue:GE has updated the Service Manuals for the LightSpeed CT and Discovery NM to update an incorrect torque value. We are asking our customers to ensure that the organization that services their equipment uses the latest version of the Service Manual to maintain the safe and proper performance of the medical device. Customers can continue to use the devices.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2016