Detail of Recall

Type of Product:Device
Medsafe Reference:20039
Brand Name:Endologix Nellix Endovascular Aneurysm Sealing System
Model:N10-100, N10-110, N10-120, N10-130, N10-140, N10-150, N10-160, N10-170,
Affected:
Software version:
Recalling Organisation:eVas Medical
73 Churchill Road
Murrays Bay
Auckland 0630
Contact Information:
Manufacturer:Endologix LLC
Issue:Updated Instructions for Use (IFU) for the Nellix EndoVascular Aneurysm Sealing System.
Feature that differentiates Nellix from other AAA endovascular devices is the use of a 2-part water soluble polymer that is used to fill the aneurysmal space. During the Nellix procedure, polymer is dispensed through fill tubes within the catheter into endobags which surround the stent grafts. The polymer is introduced in a liquid form and begins to crosslink as soon as the components are mixed. Complete crosslinking occurs inside the endobags 3 to 5 minutes after filling has been completed. The polymer will not crosslink in the fill tubes, as long as there is
continuous flow of the solution. Once the polymer stops moving through the fill tubes, crosslinking can occur in 30 seconds within the fill tubes. If the polymer cures in the fill tubes, it can impede further introduction of polymer.Utilization of the secondary fill option (outline din IFU) is necessary to safely compete the filling process.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/4/2016