Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20036 |
| Brand Name: | CAPNOSTREAM 20P |
| Model: | Capnostream 20P, Capnostream 20PX1 |
| Affected: | CS08658-01, CS08652-01 |
| Software version: | |
| Recalling Organisation: | Covidien NZ Ltd Level 3, Building 5, Central Park Corporate Centre 666 Great South Road Penrose AUCKLAND 1051 |
| Contact Information: | 0508 489 264 |
| Manufacturer: | Oridion Medical 1987 Ltd |
| Brand Name: | CAPNOSTREAM 20P NX1 |
| Model: | Capnostream 20P NX1 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Covidien NZ Ltd Level 3, Building 5, Central Park Corporate Centre 666 Great South Road Penrose AUCKLAND 1051 |
| Contact Information: | 0508 489 264 |
| Manufacturer: | |
| Issue: | Corvidien is issuing a recall for the battery pack used in Oridion labeled Capnostream™20p patient monitors. The scope of this recall includes battery packs that were manufactured between April 2014 and February 2016. The recall includes battery packs included with the monitors and spare replacement parts purchased separate from the monitor. This recall is due to a supplier defect in battery manufacturing that may increase the risk of thermal damage to the battery pack. |
| Recall action type: | Correction |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/4/2016 |