Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20028 |
| Brand Name: | UniCell DxH600 UniCell DxH800 DxH SMS |
| Model: | 629029, B24465, B24802,B68304 B23858 and 775222 |
| Affected: | |
| Software version: | DxH 800/SMS 3.0.2.0 and DxH 600 1.1.1.0 released in June 2015 |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The DxH 800, DxH 600, and DxH Slidemaker Stainer may fail to notify the user via the Alert Icins, Audible Alarm, and Instrument Beacon, for daily check failures, LIS disconnects printer failurees and reagent container low/depleted notifications. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/4/2016 |