Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20023 |
| Brand Name: | Siemens e.cam and all Symbia systems with foresight detectors |
| Model: | |
| Affected: | 10151532, 8717733, 10275009, 10275010, 10275879, 8717741, 10275008, 10275007, 10764804, 10764803, 10764801, 10764802, 10910501, 10151531 with foresight detectors |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Medical Solutions USA Inc |
| Issue: | Affected systems performing gated or dynamic acquisitions may lose some detector time-information this may result in framing of the image data and the calculated ejection fraction values may be lower. Misdiagnosis may be possible. The resulting risk is that the framing of the image data and the calculated ejection fractions values may be lower. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/4/2016 |