Type of Product:Device
Medsafe Reference:20020
Brand Name: Drager D-Vapor
Model:drager D Vapor, Drager D Vapor 300
Affected:
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue: As part of our continued market and product monitoring we have become aware of isolated cases in which the unwanted release of evaporating agent was observed by users at the end of the filling procedure. The release may lead to spraying of aerosol into face and eyes of the operator. Eye contact with volatile anaesthetics can result in serious irritation; immediate flushing of eyes and consultation of a physician is recommended. Recently updated Material Safety Data Sheets (MSDS) of Desflurane manufacturers recommend the wearing of safety glasses when handling these substances.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/4/2016