Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19990
Brand Name:Medtronic CareLink Monitor
Model:2490C
Affected:
Software version:version MR7
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:Medtronic has determined that a recent software update was distributed with incorrect telephone number information used by the home monitor to contact the CareLink Network. Monitors that have received this update may no longer be able to connect to the CareLink Network as a result of this error. This inability to connect will prevent the monitor from transmitting implanted device data, including CareAlerts, to the clinic and from receiving future software updates.
This issue does not impact home monitors currently using a cellular or internet accessory (e.g. WireX) to connect to the CareLink Network.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/4/2016
Hide menus