Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19985 |
| Brand Name: | GE Anaesthesia Work Stations |
| Model: | |
| Affected: | Multiple serial numbers |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | Manufacturer has recently become aware of a potential safety issue with the optional Vacuum Suction Function of certain Aisys family of Anaesthesia products. The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 lpm required by ISO 10079-3. If this issue is left unresolved, it could result in inadequate suction performance to clear the airway and inadequate ventilation of the patient. There have been no injuries reported as a result of this issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/4/2016 |