Type of Product:Device
Medsafe Reference:19985
Brand Name:GE Anaesthesia Work Stations
Model:
Affected:Multiple serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:Manufacturer has recently become aware of a potential safety issue with the optional Vacuum Suction Function of certain Aisys family of Anaesthesia products. The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 lpm required by ISO 10079-3. If this issue is left unresolved, it could result in inadequate suction performance to clear the airway and inadequate ventilation of the patient. There have been no injuries reported as a result of this issue.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/4/2016