Detail of Recall

Type of Product:Device
Medsafe Reference:19972
Brand Name:ADVIA Centaur XPT
Model:ADVIA Centaur XPT
Affected:10711433
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Certain reservoirs have been manufactured with walls that are below the nominal specification, potentially causing a following problem when the reservoir is under vacuum. The seam at the bottom of certain reservoirs could crack slightly and have a slight leak.
The fitting at the bottom of the reservoir could exhibit some leaks at the interface between the reservoir and the fitting.
If a leak occurs due to a crack in the bottom of the reservoir, foaming may occur in the glass jar in the bulk fluid drawer and cause the instrument to stop with vacuum low error # 01 600 13 24.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/6/2016