Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19968 |
| Brand Name: | Siemens AXIOM- AX |
| Model: | Artis zee system |
| Affected: | version VC21B |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens AG |
| Issue: | In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. b) In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator. Follow up: The manufacturer has recently become aware of an additional affected system that we were not initially aware of. The additional affected customer is Auckland Hospital. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/4/2016 |