Detail of Recall

Type of Product:Device
Medsafe Reference:19955
Brand Name:Mobile Digital X Ray System
Model:MobileDiagnost wDR Rel. 1 and 2
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The power plug, which has been delivered with the system is not according to the local regulation in New Zealand (AS/NZS 3115:2000).
The pins of the plug can be touched when the plug is not inserted completely into the mains outlet, potentially exposing the person to electrical shock.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/4/2016