Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19954
Brand Name:Cook Inc Catheter, Angiography
Model:
Affected:all lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Group Inc
Brand Name:Cook Inc Aprima Access Set
Model:
Affected:all lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Brand Name:Cook Inc Transluminal Biliary Biopsy Forceps Set
Model:
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Issue:Manufacturer is initiating a recall of listed products as they have identified an increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/ or separation.
Preliminary investigation into this matter has identified that environmental conditions such as storage temperature, humidity and the use of Vapourised Hydrogen Peroxide for whole-room decontamination within healthcare facilities, may be contributing to this occurrence.
Potential adverse events that may occur as a result if catheter polymer degradation coudl include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. (eg; device fragments in the vascular system, genitourinary system or other soft tissues).
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/4/2016
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /