Detail of Recall

Type of Product:Device
Medsafe Reference:19945
Brand Name:GE Anaesthesia Work Stations
Model:
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:It has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. Aisys CS2 anesthesia devices and upgraded Aisys anesthesia devices deliver a momentary, self-correcting increase of the anesthetic agent, affecting the inhaled and exhaled concentrations for a short time, upon either of the following setting changes: A fresh gas setting change from 95%-25% oxygen to only air (21% oxygen). Any total flow setting change while using 21% oxygen (air only). The momentary bolus of anesthetic agent will not occur with nitrous oxide as the carrier (balance) gas or if the oxygen concentration is set above 21% (air only). Delivery of this momentary bolus of anesthetic agent is potentially hazardous because it could lead to hypotension in certain vulnerable pediatric patients when 21% oxygen (air only) is used. There have been no injuries re ported of this issue.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2016