Detail of Recall

Type of Product:Device
Medsafe Reference:19942
Brand Name:Siemens Dimesnion Vista System
Model:6L119UL
Affected:All lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue: Testing with a new lot of antibody showed On-Board Stability claim of 20 hours for Dimension Vista Methotrexate Application was not met. Lots G2 and H1 (current antibody) showed similar results. Further testing on lot G2 showed 4 hours On-Board Stability.( Description of issue – Siemens Healthcare Diagnostics has confirmed that the Syva® Emit® Methotrexate Application for the Dimension Vista® System does not meet the On-Board Stability claim of 20 hours).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2016