Detail of Recall

Type of Product:Device
Medsafe Reference:19936
Brand Name:Maquet Cardiosave Hybrid and Rescue IABP
Model:FLOW-i C20 FLOWE-i C30 and FLOW-i C40
Affected:Part Nos: 66 77 200 ,66 77 300 and 66 77 400
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet Critical Care AB
Issue:The alarm TE63/64 is generated during some circumstances when the FLOW -i-Anaesthesia Systems considers that the internal communication does not work as intended. This falsely triggered technical alarm results in that the vaporiser is turned off due to burning issue between the sub systems.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2016