Detail of Recall

Type of Product:Device
Medsafe Reference:19926
Brand Name:Zimmer UniWallis System Instruments
Model:
Affected:all lots
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Zimmer Spine
Issue:Since the market launch of this product, the instruments were distributed with Instruction for Use related to the method for cleaning and disinfection and for steam sterilisation at health care facilities; however, the method was not adequately validated.
Instruments cleaned, disinfected, and sterilised under the former IFU for the UniWallis systems may not be adequately cleaned, disinfected and sterilised resulting in a risk of infection.
Manufacturer to provide updated cleaning disinfection and sterilisation instruction for use of UniWallis™ System.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/3/2016