Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19898 |
| Brand Name: | Stryker Neurovascular - III - 46352 - Neurovascular embolisation implant |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Neurovascular |
| Issue: | It was identified that the Neuroform Atlas pouch and carton label references the incorrect measurement in millimeters for the Minimum Catheter Diameter. The Minimum Catheter Diameter is correctly listed as 0.0165 inches on both product labels, but incorrectly converted to millimetres as 0.69mm. The correct conversion value is 0.42. The labelling discrepancy could lead to a physician selecting a catheter which is too large. The hazard associated with this is that the stent would prematurely deploy during transfer into the catheter. The result to the patient being extended procedure/ additional treatment required and increased time under anaesthesia. No adverse health consequences are anticipated. There is no risk to patients associated with previous use of affected devices. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/3/2016 |