Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19883 |
| Brand Name: | Ellex 2RT Laser System and Spare Eye Filters |
| Model: | LR1532 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Ellex Medical Pty Ltd 82 Gilbert Street Adelaide SA 5000 Australia |
| Contact Information: | |
| Manufacturer: | Ellex Medical Pty Ltd |
| Issue: | 11 fixed safety filters were returned from the field after motorised safety filterupgrade. On inspection 2 of the 11 were found to have a small crack along one edge. Evaluation of the optics found that the Optical Density was within specification however found that the glass in one particular area on both of them was cracked. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/9/2016 |