Detail of Recall

Type of Product:Device
Medsafe Reference:19875
Brand Name:Hamilton/47244/Ventilators
Model:
Affected:
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:The analysis of the customer complaint has identied an issue related to suctioning with the use of the Hamilton-G5 and Hamilton-S1 ventilators. After preforming the suctioning manuoeuvre, including disconnecting the patient suctioning and then reconnecting the patient the preset pattern of the ventilation may not continue as expected. This software flaw may lead to either hyperventilation or hypoventilation.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/3/2016