Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19871 |
| Brand Name: | Siemens ADVIA Chemistry Systems Triglyceride Concentrated Reagent |
| Model: | 10697575 |
| Affected: | Lot numbers: 348297 and 359931 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Manufacturer has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions for Use (IFU) linearity claim at the upper limit of the assay range as it approaches end of shelf life. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/3/2016 |





