Detail of Recall

Type of Product:Device
Medsafe Reference:19856
Brand Name:Philips Allura Xper
Model:
Affected:
Software version:FD R8.1.16
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:Manufacturer has discovered through customer complaints and internal testing that uncontrolled geometry movements can occur when the sytem is not switched on/off regularly. Due t the nature of the proble the ser will perceive a gradually increased sluggishness. Upon activation of the IU controls an uncontrolled geometry movement casn occur. The uncontroled movements are immediateky stipped upon release of the IU controls.
Recall action type:Correction
Recall action:
Level of recall:
Date Commenced:4/3/2016