Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19811
Brand Name:Roche Elecsys Estradiol 11 and 111 Assays
Model:03000079190 and 06656021190
Affected:all lot numbers
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:If the estradiol status of postmenopausal women under treatment with fulvestrant is tested with Elecsys Estradiol II or Elecsys Estradiol III assays, an interference leading to falsely increased results of Estradiol due to the drug may occur. Subsequently the incorrect level of estradiol may lead to misinterpretations of the hormone status and the use of fulvestrant may be altered. In addition the efficiency of anti-estrogen treatment might be underestimated. A medical risk for postmenopausal women under fulvestrant treatment cannot be excluded.
Recall action type:Correction
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/3/2016