Detail of Recall

Type of Product:Device
Medsafe Reference:19802
Brand Name:Siemens Dimension Blood Urea Nitrogen (BUN) Flex Reagent Cartridge
Model:K1021
Affected:Lot Nos: 15215AE 15243BB 15264BA 15299BB 15300BA 1532OBB 15341AC
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics (UK) Ltd
Issue:The manufacturer has determined that Dimension Vista® Blood Urea Nitrogen (BUN) lots 15215AE, 15243BB, 15264BA, 15299BB, 15300BA, 15320BB, and 15341AC may exhibit inaccurate patient and/or Quality Control results. The issue has been isolated to Flex® reagent cartridges manufactured from specific mold cavities. Investigation indicates loss of activity of the GLDH reagent wells from specific mold/cavity identifiers.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/2/2016