Detail of Recall

Type of Product:Device
Medsafe Reference:19795
Brand Name:Zimmer Biomet Orthopedics Verso Shoulder Forked Retractor
Model:PN 403852
Affected:all lots
Software version:
Recalling Organisation:Biomet New Zealand Ltd
Contact Information:0275 522 595
Manufacturer:Biomet Orthopedics Inc
Issue:The manufacturer has identified that specific model (PN: 402582 )was manufactured using 420 stainless instead of the material specified on the print, 420 S29 stainless steel. 420 stainless is more brittle than 420 S29.
The instrument may facture duriing use. If the retractor fractures during surgery, surgical intervention may be necessary to retrieve any fractured pieces. A delay in surgery greater than 30 minutes may occur. If the pieces are unable to be retrieved, the patient would retain a foreign body.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/2/2016