Detail of Recall

Type of Product:Device
Medsafe Reference:19794
Brand Name:Medtronic Reveal LINQ Insertable Cardiac Monitor (ICM)
Model:LNQ11
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:An issue has been identified with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the RRT alert in some devices. As of February 12, 2016, Medtronic has observed an occurrence rate of 0.45% of devices experiencing this issue. Battery capacity is not affected and the device will continue to support data collection and manual data transmissions. Medtronic is preparing a software update to correct this issue in the field, reset RRT & EOS status, and re-enable wireless transmissions as necessary.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2016