Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19794 |
| Brand Name: | Medtronic Reveal LINQ Insertable Cardiac Monitor (ICM) |
| Model: | LNQ11 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | An issue has been identified with the sensitivity of an algorithm used in the Reveal LINQ ICM that may prematurely trigger the RRT alert in some devices. As of February 12, 2016, Medtronic has observed an occurrence rate of 0.45% of devices experiencing this issue. Battery capacity is not affected and the device will continue to support data collection and manual data transmissions. Medtronic is preparing a software update to correct this issue in the field, reset RRT & EOS status, and re-enable wireless transmissions as necessary. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/6/2016 |