Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19787
Brand Name:bioMerieux VIDAS TOXO IgM
Model:30202
Affected:multiple lot numbers
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:Following a complaint from a customer related to discrepant results observed when testing heat inactivated sera, bioMérieux performed an investigation concluding that the performance of VIDAS TOXO IgM kit using heat inactivated sera (56°C for 30 minutes) is not in accordance with the performance characteristics as stated in the IFU. The investigation confirmed that the index results obtained with inactivated sera are lower compared to values obtained with fresh sera and may result in false negative results.
Recall action type:Correction
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/2/2016