Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19771 |
| Brand Name: | Xavant STIMPOD NMS410 / NMS450 Nerve Stimulator |
| Model: | XT-41000, XT-41000-NA, XT-41001, XT-41001-NA, XT-45011, |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Anatech Medical Ltd P O Box 0171 9 Pirihi Road Takahiwai Ruakaka Whangarei |
| Contact Information: | |
| Manufacturer: | Xavant Technology |
| Issue: | FSN concerning an update of the instruction for use (IFU), which includes changes to the sections warnings, and introduces instructions to the user to aid in the safe and precise use of the medical device. 1. Inspect all parts for any damage or manipulation. Never use any damaged or manipulated part! 2. If an electrically conductive surface of the Stimpod device or its cables is exposed, such electrically conductive surface may shock a person handling it. Do not use such a device or accessory, please contact the manufacturer for repair. 3. The refractory period delay is set at a default value to prevent the user from repeating stimulation while the nerve synapse is recovering from effects of the previous stimulation as per example: a refractory period of less than 12 seconds in TOF mode is not advisable as measurements might not represent the effect of blocking agents on the neuromuscular junction. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/2/2016 |