Detail of Recall

Type of Product:Device
Medsafe Reference:19771
Brand Name:Xavant STIMPOD NMS410 / NMS450 Nerve Stimulator
Model:XT-41000, XT-41000-NA, XT-41001, XT-41001-NA, XT-45011,
Affected:
Software version:
Recalling Organisation:Anatech Medical Ltd
P O Box 0171
9 Pirihi Road
Takahiwai
Ruakaka
Whangarei
Contact Information:
Manufacturer:Xavant Technology
Issue:FSN concerning an update of the instruction for use (IFU), which includes changes to the sections warnings, and introduces instructions to the user to aid in the safe and precise use of the medical device.
1. Inspect all parts for any damage or manipulation. Never use any damaged or manipulated part!
2. If an electrically conductive surface of the Stimpod device or its cables is exposed, such electrically conductive surface may shock a person handling it. Do not use such a device or accessory, please contact the manufacturer for repair.
3. The refractory period delay is set at a default value to prevent the user from repeating stimulation while the nerve synapse is recovering from effects of the previous stimulation as per example: a refractory period of less than 12 seconds in TOF mode is not advisable as measurements might not represent the effect of blocking agents on the neuromuscular junction.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/2/2016