Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 19740 |
| Brand Name: | Terumo Advanced Perfusion System |
| Model: | 801764 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Terumo Australia Pty Ltd (NZ) c/- Healthcare Logistics 56 Carrington Road Mount Albert Auckland 1025 |
| Contact Information: | 0800 66 77 57 |
| Manufacturer: | Terumo Cardiovascular Systems Corp |
| Issue: | Manufacturer has received three reports of a Terumo® Advanced Perfusion System 1 displaying a pressure reading of '999'. Terumo CVS has confirmed that in one of these cases, the user was unable to clear the pressure alarm and restart the pump. NOTE: The system displays '999' at pressures greater than +900mmHg. The reading can indicate a true high pressure or a malfunction of the pressure transducer or cable. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/2/2016 |