Detail of Recall

Type of Product:Device
Medsafe Reference:19740
Brand Name:Terumo Advanced Perfusion System
Model:801764
Affected:
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:Manufacturer has received three reports of a Terumo® Advanced Perfusion System 1 displaying a pressure reading of '999'. Terumo CVS has confirmed that in one of these cases, the user was unable to clear the pressure alarm and restart the pump.
NOTE: The system displays '999' at pressures greater than +900mmHg. The reading can indicate a true high pressure or a malfunction of the pressure transducer or cable.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/2/2016