Detail of Recall

Type of Product:Device
Medsafe Reference:19737
Brand Name:Bio-Rad Liquichek Specialty Immunoassay Controls (Level 1 to 4)
Model:359 359X 364 365 and 366
Affected:Master Lot Nos: 57430 57440 57450 57460 57470 57480 and 57490
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:Manufacturer has observed that Intact PTH may not meet the thawed unopened and thawed open vial stability claims as printed in the package inserts.
(5 customers reported Intact PTH failing to meet the 23 days thawed open vial stability claim at 2 to 8 deg C.0
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/2/2016