Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19733
Brand Name:Evalve MitraClip
Model:SGC01ST
Affected:
Software version:
Recalling Organisation:Abbott Vascular Division of Abbott Laboratories (NZ) Ltd
Ground Floor, Building D
4 Pacific Rise
Mount Wellington
Auckland 1060
Contact Information:0800 827 285
Manufacturer:Evalve Inc
Brand Name:Evalve Mitraclip Clip Delivery System (CDS)
Model:MSK02ST
Affected:
Software version:
Recalling Organisation:Abbott Vascular Division of Abbott Laboratories (NZ) Ltd
Ground Floor, Building D
4 Pacific Rise
Mount Wellington
Auckland 1060
Contact Information:0800 827 285
Manufacturer:Evalve Inc
Issue:Manufacturer has recently received reports of cases in Clip Delivery System devices that contain the One-Way Actuator Knob where a user attempted implanting a MitralClip but the Clip could not be detached from the delivery system due to a mandrel fracture. These cases resulted in severe surgical interventions and in one case the patient died post-operatively due to severe comorbidities.
Current Instructions for Use (IFU) require the Arm Positioner to be in a Neutral position prior to turning the Actuator Knob to deploy the Clip; manufacturerr is revising the MitraClip IFU Clip deployment sequence to provide additional assurance that tension is completely eliminated prior to deploying the Clip. Abbott Vascular will train all MitraClip implanters on the revised instructions.
Note : Currently, there are no MitraClip systems in New Zealand.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:5/2/2016