Detail of Recall

Type of Product:Device
Medsafe Reference:19704
Brand Name:Olympus TJF-Q180V Duodenovideoscope
Model:
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:Nature of correction.:
There will be:
1. Minor changes to parts around the forceps elevator to improve the tolerance
2. Minor changes to the Reprocessing Manual (the forceps elevator must be raised and lowered during the leakage testing and updated warnings for AER and reprocessing workflow)
3. Changes to the Operation Manual (each TJF-Q180V will be required to be returned to Olympus for an annual inspection of the forceps elevator surrounding parts)
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/2/2016