Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:19693
Brand Name:Bio-Rad D-10 Rack Loader
Model:220-0600
Affected:multiple items and batches
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Biomet France SARL
Issue:There is a potential for a sample and result to be mismatched when running in the D-10 Rack Loader configuration. This means that a patient result may be assigned to an incorrect sample ID.
While the manufacturer investigates and prepares a permanent solution, the recommendation is that customers verify that the number of results equals the total number of samples run. If there is a discrepancy, customers should identify the rack with the missing result, rerun all samples from that rack, and check the laboratory information system (LIS) results.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/1/2016