| Brand Name: | Bio-Rad D-10 Rack Loader |
| Model: | 220-0600 |
| Affected: | multiple items and batches |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Biomet France SARL |
| Issue: | There is a potential for a sample and result to be mismatched when running in the D-10 Rack Loader configuration. This means that a patient result may be assigned to an incorrect sample ID. While the manufacturer investigates and prepares a permanent solution, the recommendation is that customers verify that the number of results equals the total number of samples run. If there is a discrepancy, customers should identify the rack with the missing result, rerun all samples from that rack, and check the laboratory information system (LIS) results. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/1/2016 |