Type of Product:Device
Medsafe Reference:19673
Brand Name:Varian Eclipse Treatment Planning Program
Model:
Affected:
Software version:versions 11, 13.0, 13.5, 13.6
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Issue:Manufacturer has received a report of an anomaly involving Eclipse™ Treatment Planning System Pencil Beam Convolution (PBC) 11.0.31 to compute a conventional arc dose distribution. When using PBC 11.0.31 to calculate the dose for a conventional arc field with more than 100 segments with Eclipse versions 11, 13, 13.5 or 13.6, the displayed dose does not correspond to the calculated Monitor Units (MU). Varian has not received any report of misadministration due to this issue.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/1/2016