Detail of Recall

Type of Product:Device
Medsafe Reference:19666
Brand Name:Bio-Merieux API ZYM B reagent
Model:70493
Affected:Lot Nos: 100375610, 1003772410, 1003772470, 100387970, 1003790030, 1004077210, 1004097610, 1004355410 and 1004355730
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:BioMerieux Australia Pty Ltd
Issue:A customer reported a false negative result with Streptococcus group D strain using reagent API ZYM B. Customer retested the S,pyogenes strain on diferent lots and no colour changes were detected.
Manufacturer's investigaton confirmed a loss of performance due to light exposure after 3 months of shelf life for API ZYM B which lead to false negative results on some API strups.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/1/2016